CALDUR® Ascorbic acid – Colorimetric

Ascorbic acid determination in water using a color chart (0.1-10 mg/l C₆H₈O₆)
Article number: 
depends on selection Categories , ,
Select the desired properties:
Quantity
iHow it works
  1. 1Add to your quote
  2. 2Review & submit your list
  3. 3Personal quote – within 24 h on business days

Fast, precise on-site analysis of your water quality – reliable results without lab effort.

Made in GermanyISO 9001 (TÜV NORD)Experience since 1975personal advice
Request a Quote

Private customers can also find our range in our partner store Ampha-Analyse.

Reagent for CALDUR® Ascorbic acid – Colorimetric:
See also:

CALDUR® Ascorbic acid test kit – Colorimetric

What is ascorbic acid measured for?

Ascorbic acid (vitamin C, C₆H₈O₆) is used in water treatment as an oxygen binding and reducing agent. In the food industry, ascorbic acid is used as an antioxidant and preservative to prevent the oxidation of food and extend its shelf life.

In steam boilers and closed cooling systems, ascorbic acid is used in a similar way to Sulfite to bind dissolved oxygen in order to prevent corrosion. Ascorbic acid reacts with oxygen to form dehydroascorbic acid and thus protects pipes and system components from oxygen corrosion.

The optimum ascorbic acid concentration in technical applications is typically 10-50 mg/l C₆H₈O₆. In the food industry, the concentrations vary greatly depending on the application and must be controlled on a product-specific basis.

How does the test set work?

The CALDUR® Ascorbic Acid Test Set – Colorimetric is based on the ferrozine method. Ascorbic acid reduces iron(III) to iron(II), which then forms a violet color complex with ferrozine. After adding two reagents to the water sample, a characteristic violet color develops within 3 minutes, the intensity of which is proportional to the ascorbic acid concentration.

The measured value is determined by direct color comparison with a 6-stage color chart provided. The method is specific for reducing substances such as ascorbic acid.

Advantages of the CALDUR® test set

  • Quick analysis in just 3 minutes

    Ideal for regular checks in boiler systems and production processes – immediate results

  • Wide measuring range from 0.1 to 10 mg/l C₆H₈O₆

    Covers the entire field of practical relevance for technical and food applications

  • 100 provisions per set – economical in continuous use

    Cost-efficient for daily checks in production facilities and quality assurance

  • Colorimetric use with color chart

    Easy handling without additional devices – perfect for mobile applications

  • Specific ferrozine method

    Proven measuring principle with high reliability

  • Robust, waterproof plastic housing with hinged lid

    Safe transportation and protection of reagents – durable and reliable

Areas of application

Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment

Carrying out the measurement:

  1. Fill water sample into test tube up to the 10 ml mark
  2. Add 10 drops of reagent A and mix
  3. Add 10 drops of reagent B and mix
  4. Wait for 3 minutes reaction time (coloration develops)
  5. Evaluation by color comparison with color chart

💡 Tip for best results

For best results, the water sample should be clear and free of suspended solids.
If the water is cloudy, pre-filtration with a filter paper (e.g. Whatman No. 1) is recommended.
The reaction time of 3 minutes should be observed, as the color intensity can develop over time.

⚠️ Important notes

  • Bring reagents to room temperature before use (20-25°C)
  • Observe a response time of 3 minutes for precise measurement results
  • Avoid direct sunlight during the measurement
  • Rinse test tubes thoroughly with distilled water after use
  • Other reducing substances can influence the measurement result

Measuring range

Measuring level Description
0.1 mg/l Very low concentration (light violet)
0.25 mg/l Low concentration (violet)
0.5 mg/l Elevated concentration (pink-violet)
1.0 mg/l Medium concentration (pink-violet)
5.0 mg/l High concentration (dark violet)
10.0 mg/l Very high concentration (intense violet)
💡 Read intermediate values:
If the color is between two fields, you can form the mean value
(e.g. between 0.5 and 1.0 mg/l = approx. 0.7-0.8 mg/l).

Number of possible provisions

You can carry out 100 measurements with one CALDUR® ascorbic acid test kit.
This refers to the quantity of reagents contained.

Measuring accuracy and reproducibility

Readability: ± 1 color level
Reproducibility: Very good if the application instructions are followed
Interference: Other reducing substances (e.g. Sulfite, hydrazine) can also be measured and lead to increased measured values. Turbidity and coloration of the sample should be avoided (filter the sample if necessary).

What to do if measured values are out of range?

Measured value > 10 mg/l

Dilution Procedure Calculation of the measured value
1:1 5 ml sample + 5 ml distilled water Measured value × 2
1:4 2 ml sample + 8 ml distilled water Measured value × 5
1:9 1 ml sample + 9 ml distilled water Measured value × 10

Measured value < 0.1 mg/l

The measured value is below the detection limit. This indicates
that the ascorbic acid concentration is very low.

Specifications at a glance

Reagents (reagent A, B)

Property Value
Parameters Ascorbic acid (C₆H₈O₆)
Item number 31-072220
Method Colorimetric (ferrozine)
Measuring principle Color comparison with 6-stage color chart
Measuring range 0.1 to 10 mg/l C₆H₈O₆
Color scale 6 measuring levels (0.1 / 0.25 / 0.5 / 1.0 / 5.0 / 10.0 mg/l)
Analysis time 3 minutes
Number of provisions 100
Packaging Hinged lid box
Packaging size 103 x 91 x 36 mm
weight approx. 160 g
Scope of delivery
2 reagent bottles (reagent A, B) with detailed labeling
1 test tube with 10 ml scale
1 color card with 6 measuring levels
Detailed operating instructions (DE/EN)
Safety data sheets for all reagents
Robust hinged lid box (103 x 91 x 36 mm)

Reagent details

Reagent Volume Shelf life Notes
Reagent A 15 ml 48 months (48M) Shake before use
Reagent B 15 ml 48 months (48M) Shake before use

Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and away from direct sunlight.

Additional information

Packaging

Hinged lid box, Refill pack

Product category

Testset

Flyer
Operating instructions