CALDUR® Chlorine dioxide – Colorimetric

Determination of chlorine dioxide in water using a color chart (0.05-0.5 mg/l)
Item no.: 
depends on selection Categories , ,
Choose a version:
Quantity
iHow it works
  1. 1Add to quote
  2. 2Review & submit your list
  3. 3Personal quote – within 24 h on working days

Fast, precise on-site analysis of your water quality – reliable results without lab effort.

Made in GermanyISO 9001 (TÜV NORD)Experience since 1975Personal advice
Request a Quote

Private customers can also find our range in our partner store Ampha-Analyse.

Reagent for CALDUR® Chlorine dioxide – Colorimetric:
See also:

CALDUR® Chlorine Dioxide Test Set – Colorimetric

What is chlorine dioxide measured for?

Chlorine dioxide (ClO₂) is a highly effective disinfectant that is used in drinking water treatment, the food industry and industrial water treatment. Compared to chlorine, chlorine dioxide has the advantage that it is also effective at higher pH values and does not form chloramines that impair the taste.

Precise monitoring of the chlorine dioxide concentration is crucial, as values that are too low mean insufficient disinfection, while concentrations that are too high pose health risks and can lead to undesirable by-products.

The limit value for drinking water in Germany is 0.2 mg/l (Drinking Water Ordinance). In practice, a residual value of 0.05 – 0.1 mg/l is often aimed for in order to ensure a continuous disinfection effect in the pipe network.

How does the test set work?

The CALDUR® Chlorine Dioxide Test Set – Colorimetric works according to the proven DPD method (diethyl-p-phenylenediamine). After adding two reagents to the water sample, a characteristic pink to violet coloration develops within about 1 minute, the intensity of which is proportional to the chlorine dioxide concentration.

The measured value is determined by comparison with a supplied color chart (colorimetric). A photometric measurement with the WINLAB Data Line photometer is also possible as an option.

Advantages of the CALDUR® test set

  • Fastest analysis in just 1 minute

    Ideal for on-site measurements – enables immediate process adjustments

  • High sensitivity from 0.05 mg/l

    Perfect for compliance with the Drinking Water Ordinance and residual value monitoring

  • 140 determinations per set – economical in continuous use

    Cost-efficient for regular inspections in waterworks and industry

  • Can be used colorimetrically and optionally photometrically

    Flexible for simple on-site tests or precise laboratory analyses with photometer

  • Easy handling without special equipment

    No additional technology required – ideal for quick tests on the go

  • Robust, waterproof plastic housing with hinged lid

    Safe transportation and protection of reagents – durable and reliable

Areas of application

Drinking / tap water
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment

Carrying out the measurement:

  1. Fill water sample into test tube up to the 10 ml mark
  2. Add 5 drops of reagent A (DPD) and swirl briefly
  3. Add 3 drops of reagent B (buffer) and mix
  4. After waiting 1 minute, compare the color with the color chart
  5. Evaluation by color comparison with color chart or photometric measurement

💡 Tip for best results

For precise measurements, the water sample should be clear. Carry out the color comparison in good lighting (daylight). Avoid direct sunlight, as this can affect color perception. The reaction time of 1 minute should be strictly adhered to, as the coloration can intensify over time.

⚠️ Important notes

  • Bring reagents to room temperature before use
  • Observe waiting time of 1 minute exactly for precise measurement results
  • The simultaneous presence of free chlorine can lead to adulteration
  • Rinse test tubes thoroughly with distilled water after use

Measuring range

Measuring level Description
0.05 mg/l Very low concentration (pale pink)
0.1 mg/l Low concentration (light pink)
0.2 mg/l Average concentration (pink)
0.3 mg/l Increased concentration (strong pink)
0.4 mg/l High concentration (pink-violet)
0.5 mg/l Very high concentration (violet)
💡 Read intermediate values:
If the color lies between two fields, you can form the mean value
(e.g. between 0.1 and 0.2 mg/l = approx. 0.15 mg/l).

Number of possible determinations

You can carry out 140 measurements with a CALDUR® chlorine dioxide test kit.
This refers to the quantity of reagents contained.

Measuring accuracy and reproducibility

Readability: ± 1 color level
Reproducibility: Very good if the application instructions are followed
Interference: Free chlorine can influence the result. If chlorine and chlorine dioxide are present at the same time, the combined test set should be used. Turbidity of the sample can falsify the result (filter the sample if necessary).

What to do if measured values are out of range?

Measured value > 0.5 mg/l

Dilution Procedure Calculation of the measured value
1:1 5 ml sample + 5 ml distilled water Measured value × 2
1:4 2 ml sample + 8 ml distilled water Measured value × 5
1:9 1 ml sample + 9 ml distilled water Measured value × 10

Measured value < 0.05 mg/l

Measured value is below the detection limit. This may mean that there is insufficient disinfection. Check the dosage of your disinfection system.

Specifications at a glance

Reagents (reagent A, B)

Property Value
Parameters Chlorine dioxide (ClO₂)
Item number 31-072270
Method Colorimetric (DPD method)
Measuring principle Color comparison with 6-stage color chart
Measuring range 0.05 to 0.5 mg/l
Color scale 6 measuring levels (0.05 / 0.1 / 0.2 / 0.3 / 0.4 / 0.5 mg/l)
Analysis duration 1 minute
Number of determinations 140
Number of reagents 2 (reagent A, B)
Packaging Hinged lid box
Packaging size 103 x 91 x 36 mm
weight approx. 120 g
Scope of delivery
2 reagent bottles (reagent A, B) with detailed labeling
1 test tube with 10 ml scale
1 color card with 6 measuring levels
Detailed operating instructions (DE/EN)
Safety data sheets for all reagents
Robust hinged lid box (103 x 91 x 36 mm)

Reagent details

Reagent Volume Shelf life Notes
Reagent A (DPD) 30 ml 24 months (24M) Shake before use, store in a cool place
Reagent B (buffer) 10 ml 24 months (24M) Shake before use

Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and away from direct sunlight.

Additional information

Packaging

Hinged lid box, Refill pack

Product category

Test kit

Flyer
Operating instructions