CALDUR® Chlorine (free) – Colorimetric
Chlorine (free) determination in water by color chart (0.05-1.0 mg/l Cl₂)


Private customers can also find our range in our partner store Ampha-Analyse.
Reagent for CALDUR® Chlorine (free) – Colorimetric:
See also:
CALDUR® Chlorine (free) test set – Colorimetric
What is free chlorine measured for?
Free chlorine is the most commonly used disinfectant in the world for treating drinking water and swimming pool water. It reliably kills bacteria, viruses and other pathogens, thus ensuring hygienically safe water. In chlorination, free chlorine is the active, disinfecting form that is responsible for effective water disinfection.
Free chlorine in drinking water should be between 0.1 and 0.3 mg/l to ensure a sufficient disinfection effect. Higher concentrations are recommended in swimming pools, depending on use. Regular monitoring of the free chlorine content is essential for ensuring water quality and protecting health.
Limit values according to the Drinking Water Ordinance:
– Free chlorine: max. 0.3 mg/l Cl₂
Regular measurement of free chlorine is essential for effective disinfection. The CALDUR® Chlorine (free) test set enables the quick and precise determination of free chlorine in the low concentration range – ideal for daily routine checks in waterworks, swimming pools and industry.
How does the test set work?
The CALDUR® Chlorine (free) test set is based on the proven DPD method (N,N-diethyl-p-phenylenediamine). DPD reacts specifically with free chlorine to form a pink-violet dye, the intensity of which is proportional to the chlorine concentration.
Measurement of free chlorine: After adding the reagents, a pink coloration corresponding to the free chlorine content develops immediately. The measured value is determined by direct color comparison with a 6-stage color chart provided.
The DPD method is internationally recognized and complies with the DIN EN ISO 7393-2 standards and WHO recommendations. The colorimetric evaluation allows you to obtain precise measurement results quickly and without additional devices.
Advantages of the CALDUR® test set
- Quick analysis in just 1 minute
Immediate results for free chlorine – ideal for daily routine checks on site
- Simple determination of free chlorine
Fast and uncomplicated measurement of the active disinfection component
- Low measuring range 0.05 – 1.0 mg/l Cl₂
Perfect for drinking water monitoring and monitoring limit values
- 150 provisions per set – economical in continuous use
Cost-efficient for daily checks in waterworks, swimming pools and industry
- DPD method according to international standards
Standard-compliant according to DIN EN ISO 7393-2 – globally recognized reference method
- Robust, waterproof plastic housing with hinged lid
Safe transportation and protection of reagents – durable and reliable
Areas of application
Carrying out the measurement:
Measurement of free chlorine:
- Fill water sample into test tube up to the 10 ml mark
- Add 5 drops of reagent A and mix
- Immediately perform color comparison with color chart (within 1 minute)
Tip for best results
The measurement should be carried out immediately after sampling, as chlorine is volatile.
Avoid exposure to sunlight and do not heat the sample.
At high chlorine concentrations, the color may fade again after the maximum (bleaching) – in this case, dilute the sample.
Important notes
- Measure the sample immediately after taking it – chlorine volatilizes quickly
- Do not exceed the reaction time for free chlorine (max. 1 minute)
- Avoid direct sunlight during the measurement
- If the color fades: Dilute the sample and measure again
- Rinse test tubes thoroughly with the sample before each measurement
Measuring range
| Measuring level | Description |
|---|---|
| 0.05 mg/l | Very low concentration (light pink) |
| 0.1 mg/l | Low concentration (pink) |
| 0.25 mg/l | Normal drinking water concentration (pink) |
| 0.5 mg/l | Elevated concentration (pink-violet) |
| 0.75 mg/l | High concentration (violet) |
| 1.0 mg/l | Very high concentration (dark violet) |
If the color lies between two fields, you can form the mean value
(e.g. between 0.25 and 0.5 mg/l = approx. 0.35-0.4 mg/l).
Number of possible provisions
You can carry out 150 measurements with a CALDUR® Chlorine test set (free).
This refers to the quantity of reagents contained.
Measuring accuracy and reproducibility
Readability: ± 1 color level
Reproducibility: Very good if the application instructions are followed
Interference: High concentrations of manganese or chromium can influence the result. Ozone and bromine also react with DPD.
What to do if measured values are out of range?
Measured value > 1.0 mg/l
| Dilution | Procedure | Calculation of the measured value |
|---|---|---|
| 1:1 | 5 ml sample + 5 ml chlorine-free water | Measured value × 2 |
| 1:4 | 2 ml sample + 8 ml chlorine-free water | Measured value × 5 |
Measured value < 0.05 mg/l
Measured value is below the detection limit. A minimum concentration of 0.1 mg/l free chlorine should be present in drinking water to ensure disinfection.
Specifications at a glance
Reagents (2 reagents)
| Property | Value |
|---|---|
| Parameters | Chlorine – free (Cl₂) |
| Item number | 31-072240 |
| Method | Colorimetric (DPD) |
| Measuring principle | Color comparison with 6-stage color chart |
| Measuring range | 0.05 to 1.0 mg/l Cl₂ |
| Color scale | 6 measuring levels (0.05 / 0.1 / 0.25 / 0.5 / 0.75 / 1.0 mg/l) |
| Analysis duration | 1 minute |
| Number of provisions | 150 |
| Number of reagents | 2 |
| Packaging | Hinged lid box |
| Packaging size | 103 x 91 x 36 mm |
| weight | approx. 150 g |
| Scope of delivery |
|---|
| 2 reagent bottles (reagent A, reagent B) with detailed labeling |
| 1 test tube with 10 ml scale |
| 1 color card with 6 measuring levels |
| Detailed operating instructions (DE/EN) |
| Safety data sheets for all reagents |
| Robust hinged lid box (103 x 91 x 36 mm) |
Reagent details
| Reagent | Volume | Shelf life | Notes |
|---|---|---|---|
| Reagent A | 30 ml | 24 months (24M) | For free chlorine, shake before use |
| Reagent B | 10 ml | 24 months (24M) | Buffer, shake before use |
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and away from direct sunlight.
Additional information
| Packaging | Hinged lid box, Refill pack |
|---|---|
| Product category | Testset |