Reagent for CALDUR® Residual hardness RG 50 – Limit value:
See also:
CALDUR® Residual hardness test set RG 50 – Limit value method
What is residual hardness measured for?
Residual hardness is a decisive parameter when monitoring water softeners and
ion exchangers. After water softening, it must be ensured that the hardness formers
(calcium and magnesium) have been removed as far as possible in order to avoid limescale deposits in downstream
systems.
The RG 50 test set tests for a limit value of 0.5 °dH (5.0 mg/l CaCO₃).
This limit value is particularly relevant for heating systems.
Regular monitoring of the residual hardness is essential in order to recognize the exhaustion status of the
ion exchanger in good time and to initiate regeneration before hard water
enters the system.
How does the test set work?
The CALDUR® residual hardness RG 50 test set works according to the proven limit value method with a
color indicator. After adding a single drop of limit value indicator to the water sample,
the resulting color immediately indicates whether the residual hardness is above or below the limit value.
Green = residual hardness below 0.5 °dH – the water is sufficiently softened. Red = residual hardness above 0.5 °dH – the ion exchanger must be regenerated.
This simple yes/no determination enables an extremely quick check without time-consuming
calculations or color comparisons.
Advantages of the CALDUR® test set
Extremely fast analysis in just 1 minute
Immediate detection of whether the limit value is exceeded or not reached – ideal for frequent checks
Easy handling – only 1 drop required
No drop counting, no calculation – color shows the result directly
Limit value of 0.5 °dH
Ideal for heating systems
250 provisions per set
Particularly economical for regular routine checks
Clear color change green/red
Clear visual differentiation without room for interpretation
Only one reagent required
Minimal effort and no risk of confusion during implementation
Robust hinged lid box or compact folding box
Optimum protection of reagents – suitable for every application situation
Areas of application
Heating systems
Water heater
Ion exchanger monitoring
Softening systems
Building services
Housing industry
Craft businesses
Facility Management
Carrying out the measurement:
Fill water sample into test tube up to the 10 ml mark
Add 1 drop of limit indicator RG 50
Close the test tubes and mix well by swirling
Assess the color of the solution (can be read off immediately)
GREEN = residual hardness below limit value (≤ 0.5 °dH) – OK
RED = residual hardness above limit value (> 0.5 °dH) – regeneration required
💡 Tip for best results
Carry out the measurement at room temperature (20-25°C).
If the water is very cold, color development may be somewhat slower.
For a clear color assessment, the sample should be clear and colorless.
If the color is uncertain, a comparative measurement with known softened water is recommended.
⚠️ Important notes
Shake reagent well before use
Only use 1 drop – more will distort the result
With intermediate colors (orange/yellowish), the residual hardness is close to the limit value
Rinse test tubes thoroughly with the sample before each measurement
Strongly colored samples can make color assessment more difficult
If the result is permanently red: regenerate the ion exchanger
Limit value
Color after addition
Meaning
Measure
GREEN
Residual hardness ≤ 0.5 °dH
Water is sufficiently softened – no action required
RED
Residual hardness > 0.5 °dH
Regenerate ion exchanger or check softening system
💡 Note on the limit value method:
The test set does not give an exact concentration, but indicates whether the value is above or below the set
limit value of 0.5 °dH. For an exact measurement of the
residual hardness, please use a titration or scientific test set.
Number of possible provisions
You can carry out 250 measurements with a CALDUR® test set residual hardness RG 50.
This refers to the quantity of limit indicator contained (10 ml bottle).
Measuring accuracy and reproducibility
Limit value: 0.5 °dH (corresponds to 5.0 mg/l CaCO₃ or 0.89 °fH) Reproducibility: Very good if the application instructions are followed Interference: Strongly colored or cloudy samples can make color assessment more difficult (filter the sample if necessary). Very high salt concentrations can influence the reaction.
Further limit value test sets available
Test set
Limit value
Typical application
RG 5
0.05 °dH
High-pressure steam boiler, ultrapure water
RG 10
0.1 °dH
Medium pressure steam boiler
RG 25
0.25 °dH
Low pressure steam boiler
RG 50
0.5 °dH
Heating systems
RG 100
1.0 °dH
General softening
Specifications at a glance
Property
Value
Parameters
Residual hardness (Ca²⁺ + Mg²⁺)
Item number
31-070260 (hinged lid box) / 31-070263 (folding box)
Method
Limit value method (complexometric)
Measuring principle
Color change when limit value is exceeded (green/red)
Limit value
0.5 °dH (5.0 mg/l CaCO₃)
Analysis duration
1 minute
Number of provisions
250
Packaging
Hinged lid box or folding box
Packaging size (box)
103 x 91 x 36 mm
Packaging size (folding box)
65 x 45 x 25 mm
Scope of delivery
Component
Component description
1 reagent bottle
Limit indicator RG 50 (10 ml) with dropper insert
1 test tube
With 10 ml scale
Operating instructions
Detailed in DE/EN
Safety Data Sheet
For the reagent
Packaging
Hinged lid box (103 x 91 x 36 mm) or folding box (65 x 45 x 25 mm)
Reagent details
Reagent
Volume
Shelf life
Notes
Limit indicator RG 50
10 ml
24 months (24M)
Shake before use
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagent should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and away from direct sunlight.