Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 ---

Calibration solution conductivity 1412 μS/cm

Conductivity calibration solution with a defined standard value of 1412 µS/cm for calibrating conductivity meters.
Article number: 
31-078015 Category
Quantity
Add to Quote
Request a quote

Private customers can also find our range in our partner store Ampha-Analyse.

Reagent for Calibration solution conductivity 1412 μS/cm:
See also:

Conductivity Calibration Solution 1412 µS/cm

Ready-to-use standard solution with a defined conductivity of 1412 µS/cm for calibrating conductivity meters.

Why do you need a conductivity calibration solution?

The measuring cell of a conductivity meter is subject to ageing and contamination, which can cause the displayed values to deviate. To ensure accurate readings, the device is calibrated regularly against a solution with an exactly known conductivity – here 1412 µS/cm.

How do you calibrate with it?

The measuring cell is immersed in the standard solution and the reading on the device is adjusted to 1412 µS/cm. This solution covers the medium conductivity range (e.g. drinking and process water).

  • Defined standard value 1412 µS/cm

    Precise reference for calibrating conductivity meters

  • Universally applicable

    Suitable for all common conductivity/TDS meters

  • Regular calibration

    Ensures consistently accurate and reproducible results

  • Ready to use

    Ready for immediate use – no preparation or dilution required

  1. Rinse the measuring cell/electrode with distilled water and dab it dry carefully
  2. Pour a small amount of standard solution into a clean container (do not measure directly in the storage bottle)
  3. Immerse the measuring cell in the 1412 µS/cm solution and allow the reading to stabilise
  4. Calibrate the meter to the setpoint of 1412 µS/cm
  5. Rinse the measuring cell after calibration

Tip for best results

Conductivity is temperature-dependent – use the solution at room temperature if possible (approx. 25 °C). Discard the portion taken after use, do not pour it back into the storage bottle.

Important notes

  • Reseal the bottle immediately after use
  • Observe the best-before date – use up opened solutions promptly
  • Store in a cool, dry place away from light

Specifications at a glance

Property Value
Type Conductivity calibration solution (standard solution)
Conductivity (setpoint) 1412 µS/cm
Article number 31-078015
Application Calibration of conductivity meters
Condition Ready to use

Scope of delivery

Conductivity Calibration Solution 1412 µS/cm 1 bottle, ready to use

Additional information

Weight 0,0000 kg
Dimensions 0,0000 × 0,0000 × 0,0000 cm