Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 ---

PC 400 upstream cooler

Sample cooler for cooling down hot water samples (up to 120 °C) before online analysis with the SYCON series.
Article number: 
30-015200 Category
Quantity
Request a quote

Private customers can also find our range in our partner store Ampha-Analyse.

Reagent for PC 400 upstream cooler:
See also:

Sample Cooler PC 400

What is the sample cooler used for?

The PC 400 sample cooler is used to withdraw a water sample from a hot water line up to 120 °C. By installing it in the sampling line and connecting a cold water line, samples such as boiler water, boiler feed water and condensate can be monitored automatically and continuously for quality. To determine the water hardness, the SYCON BoB online analyser, for example, is used downstream of the sample cooler.

Advantages at a glance

  • Compact design for wall mounting in enclosed rooms
  • Optimally matched to the analysers of the SYCON series
  • Generous cooling coil (5 m) for high cooling capacity
  • High-quality, polished stainless steel (EN 1.4571)
  • Sample water up to 16 bar, cooling water up to 8 bar

Application

The PC 400 is installed in the sampling line of a hot water line (up to 120 °C); a cold water line is connected to the cooling water inlet and outlet. The hot sample water flows through the cooling coil and releases its heat to the cooling water in counterflow. The cooled sample water is then fed to the online analyser (e.g. SYCON BoB) and continuously monitored.

Installation

  1. Mount the cooler on the wall in an enclosed room (2 fastening points, spacing 54 mm)
  2. Integrate sample water inlet and outlet into the sampling line (8 mm pipe)
  3. Connect the cooling water line to the cooling water inlet and outlet (1/2 inch pipe)
  4. Open the cooling water flow and check all connections for leaks

💡 Note on operation

For sampling from hot lines, a sufficient cooling water flow must be ensured so that the sample water is reliably cooled to analysis temperature. The permissible pressures (sample water max. 16 bar, cooling water max. 8 bar) must not be exceeded.

Specifications at a glance

Property Value
Body dimensions 103 mm × 448 mm
Cooling coil 5 m × 8 mm
Weight 4.1 kg
Material Stainless steel EN 1.4571, polished
Max. sample temperature 120 °C
Nominal pressure (cooling water side) max. 8 bar
Cooling coil (sample water) max. 16 bar
Cooling water inlet/outlet Pipe, 1/2 inch outer diameter
Sample water inlet/outlet Pipe, 8 mm outer diameter
Fastening 2-point, spacing 54 mm, screw diameter 4 mm
Installation Wall mounting in enclosed rooms
Article number 30-015200

Additional information

Weight 0,0000 kg
Dimensions 0,0000 × 0,0000 × 0,0000 cm
Product category

Accessories, Intercooler

Series

SYCON

Flyer
Operating instructions

Reagent for PC 400 upstream cooler: