Private customers can also find our range in our partner store Ampha-Analyse.
Silicate (SiO₂) is a naturally occurring water constituent that comes from the dissolution of rocks and minerals. Silicate is particularly critical in boiler water technology, as it can form glass-like deposits (silicate scale) at high temperatures and pressures, which are extremely difficult to remove.
In high-pressure boilers, the silicate content must be kept particularly low, as even low concentrations can lead to deposits on turbine blades. Typical limit values are 0.02-0.5 mg/l SiO₂ depending on the operating pressure.
Silicate control is also important in drinking water treatment and in the production of ultrapure water (e.g. for the semiconductor industry).
The CALDUR® Silicate Test Set – Photometric is based on the molybdate method. After adding three reagents to the water sample, a yellow silicomolybdate complex is formed within 8 minutes, the concentration of which is precisely determined using the WINLAB Data Line photometer.
Covers the range from boiler water to raw water
Economical for regular checks
Objective, reproducible measurement results with the WINLAB Data Line
Specific for dissolved silicate
Samples must be collected in plastic vessels – glass releases silicate and falsifies the measurement. Allow boiler water samples to cool down to room temperature before measuring.
| Property | Value |
|---|---|
| Measuring range | 0.1 to 30 mg/l SiO₂ |
| Resolution | 0.1 mg/l |
| Method | Molybdate method, photometric evaluation |
You can carry out 100 measurements with one CALDUR® Silicate Photometric Test Set.
This refers to the quantity of reagents contained.
| Application | Recommended silicate content |
|---|---|
| Low pressure boiler (< 20 bar) | < 5 mg/l SiO₂ |
| Medium pressure vessel (20-60 bar) | < 1 mg/l SiO₂ |
| High pressure boiler (> 60 bar) | < 0.5 mg/l SiO₂ |
| Feed water | < 0.02 mg/l SiO₂ |
| Verduction | Execution | Calculation of the measured value |
|---|---|---|
| 1:1 | 5 ml sample + 5 ml silicate-free water | Measured value × 2 |
Measured value is below the detection limit. For more sensitive measurements, please contact our technical support.
| Property | Value |
|---|---|
| Parameters | Silicate (SiO₂) |
| Item number | 31-073522 |
| Method | Photometric (molybdate) |
| Measuring principle | Photometric evaluation with WINLAB Data Line |
| Measuring range | 0.1 to 30 mg/l SiO₂ |
| Resolution | 0.1 mg/l |
| Duration of analysis | 8 minutes |
| Number of provisions | 100 |
| Number of reagents | 3 |
| Packaging | Plastic bag |
| Scope of delivery |
|---|
| 3 reagent bottles (reagent A, reagent B, reagent C) with detailed labeling |
| 1 round cuvette (10 ml) |
| Detailed operating instructions (DE/EN) |
| Safety data sheets for all reagents |
| Resealable plastic bag |
| Reagent | Volume | Shelf life | Notes |
|---|---|---|---|
| Reagent A (molybdate) | 30 ml | 24 months (24M) | Molybdate reagent, shake before use |
| Reagent B (acid) | 10 ml | 24 months (24M) | Acid reagent, dose carefully |
| Reagent C (reducing agent) | 10 ml | 24 months (24M) | Shake before use |
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.
| Packaging | Plastic bag, Refill pack |
|---|