Private customers can also find our range in our partner store Ampha-Analyse.
Silicate (SiO₂) enters the water naturally through the leaching of rocks and soils. Silicate control is particularly critical in boiler water and steam generator treatment: even low concentrations can lead to glass-like deposits on turbine blades and heat exchangers.
In cooling water systems, silicate in combination with calcium and magnesium promotes the formation of deposits that are difficult to dissolve. Regular checks help to avoid costly malfunctions.
Silicate plays a subordinate role in drinking water treatment, but can lead to blockages in membrane filtration processes (reverse osmosis) if the concentration is too high.
The CALDUR® silicate test set is based on the silicomolybdate blue method. Silicate reacts with molybdate in an acidic solution to form yellow silicomolybdate, which is then reduced to a blue complex. The blue coloration is proportional to the silicate concentration.
The measured value is determined by direct color comparison with a 6-stage color chart supplied.
Covers the practical range for boiler water and industry
Economical for regular checks
Immediate results on site
No measuring devices required
Safe transportation and protection of reagents
For best results, the water sample should be clear and free of suspended solids.
If the water is cloudy, pre-filtration with a filter paper is recommended.
Reagent C (oxalic acid) must be added as it suppresses phosphate interference.
| Measuring level | Description |
|---|---|
| 0.25 mg/l | Very low concentration (weak blue) |
| 0.5 mg/l | Low concentration (light blue) |
| 1.0 mg/l | Slightly increased concentration (blue) |
| 2.0 mg/l | Medium concentration (strong blue) |
| 5.0 mg/l | High concentration (dark blue) |
| 10.0 mg/l | Very high concentration (deep blue) |
You can carry out 100 measurements with one CALDUR® silicate test kit.
This refers to the quantity of reagents contained.
Readability: ± 1 color level
Reproducibility: Very good if the application instructions are followed
Interference: Phosphate in high concentrations (> 50 mg/l) can slightly increase the measured value. The addition of reagent C (oxalic acid) suppresses the phosphate interference.
| Dilution | Procedure | Calculation of the measured value |
|---|---|---|
| 1:4 | 2 ml sample + 8 ml silicate-free water | Measured value × 5 |
| 1:9 | 1 ml sample + 9 ml silicate-free water | Measured value × 10 |
measured value is below the detection limit of this test kit. This is generally not critical.
| Property | Value |
|---|---|
| Parameters | Silicate (SiO₂) |
| Item number | 31-072520 |
| Method | Colorimetric (silicomolybdate blue) |
| Measuring principle | Color comparison with 6-stage color chart |
| Measuring range | 0.25 to 10 mg/l SiO₂ |
| Color scale | 6 measuring levels (0.25 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 mg/l) |
| Analysis duration | 1 minute |
| Number of provisions | 100 |
| Number of reagents | 3 |
| Packaging | Hinged lid box |
| Packaging size | 103 x 91 x 36 mm |
| weight | approx. 170 g |
| Scope of delivery |
|---|
| 3 reagent bottles (reagent A, B, C) with detailed labeling |
| 1 test tube with 10 ml scale |
| 1 color card with 6 measuring levels |
| Detailed operating instructions (DE/EN) |
| Safety data sheets for all reagents |
| Robust hinged lid box (103 x 91 x 36 mm) |
| Reagent | Volume | Shelf life | Notes |
|---|---|---|---|
| Reagent A (molybdate) | 15 ml | 24 months (24M) | Shake before use |
| Reagent B (reducing agent) | 15 ml | 24 months (24M) | Shake before use |
| Reagent C (oxalic acid) | 15 ml | 24 months (24M) | Suppresses phosphate interference |
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25 °C) and away from direct sunlight.
| Packaging | Hinged lid box, Refill pack |
|---|