Reagent for CALDUR® Sulfide – Colorimetric:
See also:
CALDUR® Sulfide Test Set – Colorimetric
What is sulfide measured for?
Sulphide (S²-) is produced by the anaerobic decomposition of organic substances containing sulphur and can cause an unpleasant smell of rotten eggs as hydrogen sulphide (H₂S). Even low concentrations are corrosive and can damage pipelines and structures.
Sulphide is harmful to health in higher concentrations and can cause respiratory and eye irritation in people who work with contaminated water. Sulphide is undesirable in drinking water treatment and must be reliably detected.
In industrial water treatment, in sewage treatment plants and in groundwater samples, sulphide control is essential to prevent corrosion and ensure water quality.
How does the test set work?
The CALDUR® Sulfide Test Set – Colorimetric is based on the DPD method. After adding two reagents to the water sample, a characteristic coloration develops within 7 minutes, the intensity of which is proportional to the sulfide concentration.
Advantages of the CALDUR® test set
Sensitive measuring range 0.05 – 1.0 mg/l S²-
Detection of even low sulphide concentrations
100 provisions per set
Economical for regular checks
Easy handling
Fast on-site determination without measuring devices
Proven DPD method
Reliable and reproducible results
Areas of application
Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment
Carrying out the measurement:
Fill water sample into test tubes up to the 10 ml mark
Add 5 drops of reagent A and mix
Add 3 drops of reagent B and mix
Wait for 7 minutes reaction time
Perform color comparison with color chart
Tip for best results
The sample should be measured immediately after collection, as sulphide forms volatile hydrogen sulphide (H₂S). Do not shake or aerate the sample in order to avoid sulphide losses.
If H₂S can be detected by smell (from approx. 0.05 mg/l), a measurement is strongly recommended.
Important notes
Do not aerate or shake the sample – sulphide volatilizes as H₂S
Adhere exactly to the reaction time of 7 minutes
Avoid direct sunlight during the measurement
Reading may be more difficult with heavily colored samples
Thoroughly rinse the test tube with the sample before each measurement
H₂S is toxic – work in a fume cupboard at high concentrations
Measuring range
Measuring level
Description
0.05 mg/l
Very low concentration (pale yellow)
0.1 mg/l
Low concentration (light yellow)
0.2 mg/l
Slightly increased concentration (yellow)
0.3 mg/l
Medium concentration (strong yellow)
0.5 mg/l
Increased concentration (yellow-brown)
1.0 mg/l
High concentration (brown)
Read intermediate values:
If the color lies between two fields, you can form the mean value
(e.g. between 0.2 and 0.3 mg/l = approx. 0.25 mg/l).
Number of possible destinations
You can carry out 100 measurements with one CALDUR® Test Set Sulfide.
This refers to the quantity of reagents contained.
Measuring accuracy and reproducibility
Readability: ± 1 color level Reproducibility: Very good if the application instructions are followed Interference: Strongly reducing substances can influence the result. In the case of strongly colored samples (e.g. due to humic substances), the color reading may be more difficult. In these cases a photometric determination is recommended.
What to do if measured values are out of range?
Measured value > 1.0 mg/l
Verduction
Execution
Calculation of the measured value
1:1
5 ml sample + 5 ml sulfide-free water
Measured value × 2
1:4
2 ml sample + 8 ml sulfide-free water
Measured value × 5
Measured value < 0.05 mg/l
The measured value is below the detection limit. For more sensitive measurements, the photometric CALDUR® Sulfide test set (0.05-1.0 mg/l with higher resolution) can be used.
2 reagent bottles (reagent A, reagent B) with detailed labeling
1 test tube with 10 ml scale
1 color card with 6 measuring levels
Detailed operating instructions (DE/EN)
Safety data sheets for all reagents
Robust hinged lid box (103 x 91 x 36 mm)
Reagent details
Reagent
Volume
Shelf life
Notes
Reagent A (DPD 1)
30 ml
24 months (24M)
DPD reagent, shake before use
Reagent B (buffer)
10 ml
24 months (24M)
Buffer solution, shake before use
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.