Private customers can also find our range in our partner store Ampha-Analyse.
Manganese (Mn) is a naturally occurring heavy metal that is frequently found in groundwater and surface waters. The limit value for drinking water is 0.05 mg/l Mn (Drinking Water Ordinance).
Elevated manganese concentrations lead to black-brown deposits in pipes, discoloration of the water and a metallic taste. In waterworks, manganese must be reliably removed, as even low concentrations can lead to operational disruptions.
Manganese can cause deposits and corrosion in cooling water systems and industrial treatment. Regular monitoring is therefore essential.
The CALDUR® Manganese Test Set – Colorimetric is based on the formaldoxime method. After adding three reagents to the water sample, a colored complex is formed within 10 minutes, the intensity of which is proportional to the manganese concentration.
The measured value is determined by direct color comparison with a supplied color chart.
Already detects concentrations below the drinking water limit value
Economical for regular checks
Selective for manganese, low interference
Fast on-site determination without measuring devices
Safe transportation and protection of reagents
The sample should be measured immediately after collection, as dissolved manganese(II) can be oxidized to manganese(IV) on contact with atmospheric oxygen and then precipitates as brown manganese oxide. Do not filter Truebe samples before measurement if the total manganese content is to be determined.
| Measuring level | Description |
|---|---|
| 0.03 mg/l | Very low concentration (pale yellow) |
| 0.05 mg/l | Drinking water limit value (light yellow) |
| 0.1 mg/l | Increased concentration (yellow) |
| 0.2 mg/l | Significantly increased concentration (strong yellow) |
| 0.5 mg/l | High concentration (yellow-orange) |
| 1.0 mg/l | Very high concentration (orange) |
You can carry out 100 measurements with one CALDUR® Manganese Test Set.
This refers to the quantity of reagents contained.
Readability: ± 1 color level
Reproducibility: Very good if the application instructions are followed
Interference: Iron in high concentrations (>5 mg/l) can influence the result. If samples are heavily colored (e.g. by humic substances), the color reading may be more difficult. In these cases, a photometric determination is recommended.
| Verduction | Execution | Calculation of the measured value |
|---|---|---|
| 1:1 | 5 ml sample + 5 ml manganese-free water | Measured value × 2 |
| 1:4 | 2 ml sample + 8 ml manganese-free water | Measured value × 5 |
The measured value is below the detection limit. This is a good result for drinking water, as the limit value of 0.05 mg/l is safely adhered to.
Reagents (3 reagents)
| Property | Value |
|---|---|
| Parameters | Manganese (Mn) |
| Item number | 31-072360 |
| Method | Colorimetric (formaldoxime) |
| Measuring principle | Color comparison with 6-stage color chart |
| Measuring range | 0.03 to 1.0 mg/l Mn |
| Color scale | 6 measuring levels (0.03 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 mg/l) |
| Duration of analysis | 10 minutes |
| Number of provisions | 100 |
| Number of reagents | 3 |
| Packaging | Hinged lid box |
| Packaging size | 103 x 91 x 36 mm |
| weight | approx. 170 g |
| Scope of delivery |
|---|
| 3 reagent bottles (reagent A, reagent B, reagent C) with detailed labeling |
| 1 test tube with 10 ml scale |
| 1 color card with 6 measuring levels |
| Detailed operating instructions (DE/EN) |
| Safety data sheets for all reagents |
| Robust hinged lid box (103 x 91 x 36 mm) |
| Reagent | Volume | Shelf life | Notes |
|---|---|---|---|
| Reagent A (Formaldoxime) | 30 ml | 24 months (24M) | Formaldoxime reagent, shake before use |
| Reagent B (buffer) | 15 ml | 24 months (24M) | Alkaline buffer solution |
| Reagent C (oxidizing agent) | 10 ml | 24 months (24M) | Oxidizing reagent, shake before use |
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.
| Packaging | Hinged lid box, Refill pack |
|---|