Reagent for CALDUR® Aluminium – Colorimetric:
See also:
CALDUR® Aluminum Test Set – Colorimetric
What is aluminum measured for?
Aluminum (Al) enters drinking water primarily via aluminum-based flocculants (e.g. aluminum sulfate, polyaluminum chloride). These agents are used in water treatment to remove impurities. The limit value for drinking water is 0.2 mg/l Al (Drinking Water Ordinance).
Elevated aluminum concentrations can lead to turbidity and are an indication of insufficient flocculation or filtration. In acidic waters (pH < 5), natural aluminum can be dissolved from the soil and reach concentrations that are toxic to fish.
Regular monitoring of the aluminum concentration is particularly important in waterworks, in industrial treatment and in environmental analysis.
How does the test set work?
The CALDUR® Aluminium Test Set – Colorimetric is based on a colorimetric color card method. After adding three reagents to the water sample, a characteristic coloration develops within 8 minutes, the intensity of which is proportional to the aluminum concentration.
The measured value is determined by direct color comparison with a supplied color chart.
Advantages of the CALDUR® test set
Sensitive measuring range 0.05 – 2 mg/l Al
Covers the drinking water limit and above
150 provisions per set
Cost-efficient for continuous use in waterworks
Easy handling with color card
Fast on-site determination without measuring devices
Proven colorimetric method
Reliable and reproducible results
Robust plastic housing with hinged lid
Safe transportation and protection of reagents
Areas of application
Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment
Carrying out the measurement:
Fill water sample into test tubes up to the 10 ml mark
Add 5 drops of reagent A and mix
Add 5 drops of reagent B and mix
Add 3 drops of reagent C and mix
Wait for 8 minutes reaction time
Perform color comparison with color chart
Tip for best results
The sample should have a pH value between 4 and 6, as aluminum is best detectable in this range. At higher pH values, aluminum can precipitate as hydroxide and is then not detected. Filter Truebe samples before measurement.
Important notes
Adhere exactly to the response time of 8 minutes
Avoid direct sunlight during the measurement
Fluoride-containing samples can lead to lower results
Reading may be more difficult with heavily colored samples
Thoroughly rinse the test tube with the sample before each measurement
Use glass appliances – aluminum can dissolve out of plastic
Measuring range
Measuring level
Description
0.05 mg/l
Very low concentration (pale pink)
0.1 mg/l
Low concentration (light pink)
0.2 mg/l
Drinking water limit value (pink)
0.5 mg/l
Increased concentration (deep pink)
1.0 mg/l
High concentration (pink-violet)
2.0 mg/l
Very high concentration (violet)
Read intermediate values:
If the color lies between two fields, you can form the mean value
(e.g. between 0.1 and 0.2 mg/l = approx. 0.15 mg/l).
Number of possible destinations
With a CALDUR® aluminum test set you can carry out 150 measurements.
This refers to the quantity of reagents contained.
Measuring accuracy and reproducibility
Readability: ± 1 color level Reproducibility: Very good if the application instructions are followed Interference: Fluoride can interfere with the measurement as it forms stable complexes with aluminum. Iron(III) in high concentrations can also interfere. A photometric determination is recommended for heavily colored samples.
What to do if measured values are out of range?
Measured value > 2.0 mg/l
Verduction
Execution
Calculation of the measured value
1:1
5 ml sample + 5 ml aluminum-free water
Measured value × 2
1:4
2 ml sample + 8 ml aluminum-free water
Measured value × 5
Measured value < 0.05 mg/l
The measured value is below the detection limit. This is a good result for drinking water, as the limit value of 0.2 mg/l is adhered to.
3 reagent bottles (reagent A, reagent B, reagent C) with detailed labeling
1 test tube with 10 ml scale
1 color card with 6 measuring levels
Detailed operating instructions (DE/EN)
Safety data sheets for all reagents
Robust hinged lid box (103 x 91 x 36 mm)
Reagent details
Reagent
Volume
Shelf life
Notes
Reagent A (buffer)
30 ml
24 months (24M)
Buffer solution, shake before use
Reagent B (color reagent)
15 ml
24 months (24M)
Color reagent, shake before use
Reagent C (compensation)
10 ml
24 months (24M)
Compensation reagent
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.