Private customers can also find our range in our partner store Ampha-Analyse.
Reagent for CALDUR® Manganese – Photometric:
See also:
CALDUR® Manganese Test Set – Photometric
What is manganese measured for?
Manganese (Mn) occurs naturally in groundwater and at high concentrations can lead to black deposits in pipes, discoloration and an unpleasant taste. The limit value for drinking water is 0.05 mg/l Mn (Drinking Water Ordinance).
In water treatment, manganese must be reliably monitored as it can pollute the filter systems. Photometric measurement enables precise determination over a wide concentration range.
How does the test set work?
The CALDUR® Manganese test set is based on the formaldoxime method. After adding three reagents to the water sample, an orange-red coloration develops within 10 minutes, the intensity of which is precisely determined with the WINLAB Data Line photometer.
Advantages of the CALDUR® test set
Large measuring range 0.05 – 20 mg/l Mn
From the drinking water limit to high industrial concentrations
100 provisions per set
Economical for regular checks
High precision due to photometric evaluation
Objective, reproducible measurement results with the WINLAB Data Line
Proven formaldoxime method
Specific for manganese, little interference
Areas of application
Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment
Carrying out the measurement:
Fill water sample into round cuvette up to the mark
Add reagent A and mix
Add reagent B and mix
Add reagent C and mix
Wait for 10 minutes reaction time
Insert the cuvette into the WINLAB Data Line photometer and start the measurement
Important notes
Adhere exactly to the reaction time of 10 minutes
Sample must not be turbid – filter beforehand if necessary
Wipe the cuvette with a lint-free cloth before measuring
Iron in high concentrations can influence the result
Requires WINLAB Data Line Photometer (sold separately)
Measuring range
Property
Value
Measuring range
0.05 – 20 mg/l Mn
Resolution
0.01 mg/l
Number of possible destinations
You can carry out 100 measurements with a CALDUR® Manganese Photometric Test Set.
Specifications at a glance
Property
Value
Parameters
Manganese (Mn)
Item number
31-073362
Method
Photometric (formaldoxime)
Measuring principle
Photometric measurement with WINLAB Data Line Photometer