Private customers can also find our range in our partner store Ampha-Analyse.
Molybdaen (Mo) is used in water treatment as a corrosion inhibitor, especially in closed cooling circuits and heating systems. The correct dosage is crucial for effective corrosion protection while at the same time avoiding overdosing.
Depending on the type of system, typical target values are between 5 and 50 mg/l Mo. Photometric measurement enables precise monitoring of the dosage.
The CALDUR® Molybdaen test set is based on the thioglycolic acid method. The reagent forms a color complex with molybdaene, the concentration of which is precisely determined with the WINLAB Data Line photometer.
Covers the entire practical dosing range
Economical for regular checks
Objective, reproducible measurement results with the WINLAB Data Line
Quick and easy to carry out in 4 minutes
| Property | Value |
|---|---|
| Measuring range | 2.5 – 100 mg/l Mo |
| Resolution | 0.1 mg/l |
With a CALDUR® Molybdaen Photometric test kit you can carry out 120 measurements.
| Property | Value |
|---|---|
| Parameters | Molybdaen (Mo) |
| Item number | 31-073672 |
| Method | Photometric (thioglycolic acid) |
| Measuring principle | Photometric measurement with WINLAB Data Line Photometer |
| Measuring range | 2.5 to 100 mg/l Mo |
| Analysis time | 4 minutes |
| Number of provisions | 120 |
| Number of reagents | 1 |
| Packaging | Plastic bag |
| Packaging | Plastic bag, Refill pack |
|---|