Reagent for CALDUR® Phosphate HR – Photometric:
See also:
CALDUR® Phosphate HR Test Set – Photometric
What is phosphate HR measured for?
In many industrial applications, phosphates are used in significantly higher concentrations than in drinking water. In boiler water, the typical phosphate target value is 10-30 mg/l PO₄³-, depending on the operating pressure and type of system.
Phosphate acts as a hardness stabilizer in boiler water – it binds calcium and magnesium and thus prevents the formation of scale. Too low a dosage leads to deposits, too high a dosage leads to increased sludge formation.
Phosphates are used as corrosion inhibitors in cooling water systems in concentrations of 5-50 mg/l. Regular monitoring ensures optimum dosing.
How does the test set work?
The CALDUR® Phosphate HR Test Set – Photometric is based on the molybdate method. After adding a reagent to the water sample, a yellow color complex (phosphorus molybdate) is formed within 1 minute, the concentration of which is precisely determined with the WINLAB Data Line photometer.
Advantages of the CALDUR® test set
Wide measuring range 2.5 – 100 mg/l PO₄³-
Ideal for boiler water and industrial applications
100 provisions per set
Economical for regular checks
Fast response time of just 1 minute
Immediate results for process control
Proven molybdate method
Reliable and specific for orthophosphate
Areas of application
Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment
Carrying out the measurement:
Fill the round cuvette with sample up to the 10 ml mark
Add 5 drops of reagent A and mix
Wait for 1 minute reaction time
Insert the cuvette into the WINLAB Data Line photometer
Start measurement – result is displayed directly in mg/l PO₄³-
Tip for best results
Boiler water samples must be cooled to room temperature before measurement. Hot samples falsify the result. The sample should be clear and free of suspended solids.
Important notes
Allow sample to cool to room temperature before measuring
Adhere exactly to the reaction time of 1 minute
Wipe the cuvette with a lint-free cloth before measuring
Use the LR test kit for concentrations below 2.5 mg/l
Only orthophosphate is recorded
Requires WINLAB Data Line photometer (Art. 37-000100)
Measuring range
Property
Value
Measuring range
2.5 to 100 mg/l PO₄³-
Dissolution
0.5 mg/l
Method
Molybdate method, photometric evaluation
Advantage of photometric measurement:
In contrast to the colorimetric method (color chart), the photometer provides exact numerical values with a higher resolution. Intermediate values are calculated automatically – ideal for documentation and quality assurance.
Number of possible destinations
With one CALDUR® Test Set Phosphate HR Photometric you can carry out 100 measurements.
This refers to the quantity of reagents contained.
Recommended dosing ranges
Application
Recommended phosphate content
Low pressure boiler (< 20 bar)
20 – 50 mg/l PO₄³-
Medium pressure vessel (20-60 bar)
10 – 30 mg/l PO₄³-
Cooling water
5 – 20 mg/l PO₄³-
Note:
The optimum dosage depends on the type of system and the operating conditions. Consult the water chemist or the dosing system documentation for system-specific target values.
What to do if measured values are out of range?
Measured value > 100 mg/l
Verduction
Execution
Calculation of the measured value
1:1
5 ml sample + 5 ml phosphate-free water
Measured value × 2
Measured value < 2.5 mg/l
Use the CALDUR® Phosphate LR test set (31-073412) for lower concentrations from 0.4 mg/l.
Specifications at a glance
Property
Value
Parameters
Phosphate (PO₄³-)
Item number
31-073417
Method
Photometric (molybdate)
Measuring principle
Photometric evaluation with WINLAB Data Line
Measuring range
2.5 to 100 mg/l PO₄³-
Dissolution
0.5 mg/l
Analysis duration
1 minute
Number of provisions
100
Number of reagents
1
Packaging
Plastic bag
Scope of delivery
1 reagent bottle (reagent A) with detailed labeling
1 round cuvette (10 ml)
Detailed operating instructions (DE/EN)
Safety data sheets for all reagents
Resealable plastic bag
Reagent details
Reagent
Volume
Shelf life
Notes
Reagent A (molybdate)
30 ml
24 months (24M)
Molybdate reagent, shake before use
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.
Required accessories
For the photometric measurement you need a WINLAB Data Line photometer with the corresponding method for phosphate HR.
The photometer is not included in the test set and is available separately.