CALDUR® Phosphate LR – Photometric
Phosphate determination in the water using a photometer (0.4-5 mg/l PO₄³-)

Private customers can also find our range in our partner store Ampha-Analyse.
Reagent for CALDUR® Phosphate LR – Photometric:
See also:
CALDUR® Phosphate LR Test Set – Photometric
What is phosphate measured for?
Phosphate (PO₄³-) is a key nutrient in water bodies and an important control parameter in water treatment. Elevated phosphate concentrations promote algae growth and can lead to eutrophication of surface waters.
In drinking water treatment, phosphates are specifically dosed as corrosion inhibitors to protect pipelines from corrosion. The correct dosage must be monitored to ensure the desired protective effect.
Phosphate is used in cooling water systems and boiler water for hardness stabilization and as corrosion protection. Regular monitoring is essential for economical operation.
How does the test set work?
The CALDUR® Phosphate LR Test Set – Photometric is based on the molybdate method. After adding two reagents to the water sample, a blue color complex (molybdenum blue) is formed within 5 minutes, the concentration of which is precisely determined using the WINLAB Data Line photometer.
Advantages of the CALDUR® test set
- Sensitive measuring range 0.4 – 5 mg/l PO₄³-
Ideal for drinking water and low-dose applications
- 200 determinations per set
Cost-efficient for continuous use
- High precision due to photometric evaluation
Objective, reproducible measurement results with the WINLAB Data Line
- Proven molybdate method
Reliable and specific for orthophosphate
Areas of application
⚠️ Hazard information (from the label)

Phosphate LR reagent A – Danger
H290 May be corrosive to metals. H314 Causes severe skin burns and eye damage.
Full information on all components: see the safety data sheet (SDS).
Carrying out the measurement:
- Fill the round cuvette with sample up to the 10 ml mark
- Add 5 drops of reagent A and mix
- Add 5 drops of reagent B and mix
- Wait for 5 minutes reaction time
- Insert the cuvette into the WINLAB Data Line photometer
- Start measurement – result is displayed directly in mg/l PO₄³-
Tip for best results
The sample should be clear and free of suspended solids. Only dissolved orthophosphate is detected. A prior digestion is required for the determination of total phosphate.
Important notes
- Adhere exactly to the reaction time of 5 minutes
- Avoid direct sunlight during the measurement
- Wipe the cuvette with a lint-free cloth before measuring
- Thoroughly rinse the test tube with the sample before each measurement
- Use the HR test kit for concentrations above 5 mg/l
- Requires WINLAB Data Line photometer (Art. 37-000100)
Measuring range
| Property | Value |
|---|---|
| Measuring range | 0.4 to 5 mg/l PO₄³- |
| Resolution | 0.1 mg/l |
| Method | Molybdate method, photometric evaluation |
In contrast to the colorimetric method (color chart), the photometer provides exact numerical values with a higher resolution. Intermediate values are calculated automatically – ideal for documentation and quality assurance.
Number of possible determinations
You can carry out 200 measurements with one CALDUR® Phosphate LR Photometric test kit.
This refers to the quantity of reagents contained.
Measuring accuracy and reproducibility
Resolution: 0.1 mg/l
Reproducibility: Very good if the application instructions are followed
Interference: Silicate in high concentrations (> 10 mg/l SiO₂) can influence the result. Heavily colored samples can interfere with the measurement. Only orthophosphate is recorded.
What to do if measured values are out of range?
Measured value > 5 mg/l
Use the CALDUR® Phosphate HR test set (31-073417) for higher concentrations up to 100 mg/l.
Measured value < 0.4 mg/l
Measured value is below the detection limit. For more sensitive measurements, please contact our technical support.
Specifications at a glance
| Property | Value |
|---|---|
| Parameters | Phosphate (PO₄³-) |
| Item number | 31-073412 |
| Method | Photometric (molybdate) |
| Measuring principle | Photometric evaluation with WINLAB Data Line |
| Measuring range | 0.4 to 5 mg/l PO₄³- |
| Resolution | 0.1 mg/l |
| Duration of analysis | 5 minutes |
| Number of determinations | 200 |
| Number of reagents | 2 |
| Packaging | Plastic bag |
| Scope of delivery |
|---|
| 2 reagent bottles (reagent A, reagent B) with detailed labeling |
| 1 round cuvette (10 ml) |
| Detailed operating instructions (DE/EN) |
| Safety data sheets for all reagents |
| Resealable plastic bag |
Reagent details
| Reagent | Volume | Shelf life | Notes |
|---|---|---|---|
| Reagent A (molybdate) | 30 ml | 24 months (24M) | Molybdate reagent, shake before use |
| Reagent B (reducing agent) | 30 ml | 24 months (24M) | Shake before use |
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 12 months.
Store the test kit in a dry, cool place (10-25°C) and protected from direct sunlight.
Required accessories
The photometer is not included in the test set and is available separately.
Additional information
| Packaging | Plastic bag, Refill pack |
|---|---|
| Product category | Test kit |
Flyer
Operating instructions
Similar articles:
Related products
-
CALDUR® Chlorine (free) – Colorimetric
Chlorine (free) determination in water by color chart (0.05-1.0 mg/l Cl₂) -
CALDUR® Nitrate – Photometric
Determination of nitrate in water using a photometer (0.5-50 mg/l NO₃-) -
CALDUR® Sulfate 10 mg – Titrimetric
Determination of sulphate in water by precipitation titration (10 mg/l SO₄²- per drop)