What is carbonate hardness measured for?
Carbonate hardness (temporary hardness) is an important parameter in water chemistry and provides information about the concentration of hydrogen carbonate and carbonate ions in the water. It is part of the total hardness and plays a decisive role in the buffering capacity of the water (pH stability).
Monitoring carbonate hardness is essential in boiler systems, as carbonate hardness precipitates and forms scale when heated. This impairs heat transfer and can lead to serious damage to the system. The carbonate hardness in boiler water should be as low as possible (< 0.5 °dH).
The Scientific series from CALDUR® offers maximum precision and user-friendliness thanks to the innovative syringe titration. Intermediate values can be read directly on the scale without having to count drops.
How does the test set work?
The CALDUR® Carbonate Hardness Scientific test set uses a precise dosing syringe for titration. Two reagents are added to the water sample, then the titration solution is added with the syringe. The measured value is read directly from the syringe scale – no drop counting required!
This method enables a continuous reading in the range from 0 to 20 °dH and offers significantly higher precision than conventional drop titrations. The resolution is at least 0.5 °dH, whereby intermediate values can also be read without any problems.
Advantages of the CALDUR® test set
Quick analysis in just 1 minute
Ideal for regular checks – faster than conventional drip titrations
Highest precision through syringe titration
Stepless reading enables intermediate values – no drop counting required
Wide measuring range from 0 to 20 °dH
Covers all practical areas from boiler water to raw water
Approx. 200 provisions per set – economical in continuous use
Particularly economical due to high number of possible measurements
Convenient handling with dosing syringe
Ergonomic design for fatigue-free working
Robust plastic housing
Safe transportation and protection of components
Areas of application
Pond / lake
Groundwater / well water
Drinking / tap water
Aquaristics
Pool / Swimming pool
Industrial processing
Boiler water
Food preparation
Waste water treatment
Carrying out the measurement:
Fill water sample into test tube
Add reagent 1 and reagent 2 according to the instructions and mix
Fill the dosing syringe with titration solution and note the initial value
Slowly add titration solution while swirling the sample
Stop adding when the color changes from green to red/orange
Read the measured value directly from the syringe scale
💡 Tip for best results
For best results, the water sample should be clear and free of suspended solids.
The color change from green to red/orange is clearly visible.
The syringe method enables very precise readings without drop counting.
Carry out the titration slowly in order to hit the end point exactly.
⚠️ Important notes
Bring reagents to room temperature before use
Rinse dosing syringe with distilled water before and after use
Carry out titration slowly to avoid over-titration
View the color change against a white background
Clean all components thoroughly after use
Measuring range
Measuring range
Resolution
Sample quantity
0 to 20 °dH
Infinitely variable (min. 0.5 °dH)
Variable sample quantity
💡 Read intermediate values:
The syringe scale enables continuous reading. You can read precise intermediate values directly,
without being limited to integer drops.
Number of possible provisions
With a CALDUR® Carbonate Hardness Scientific test set you can carry out approx. 200 measurements.
This refers to the quantity of reagents contained.
Measuring accuracy and reproducibility
Resolution: Infinitely variable, at least 0.5 °dH Reproducibility: Excellent thanks to precise dosing syringe Interferences: No significant interferences for typical water samples
What to do if measured values are out of range?
Measured value > 20 °dH
Dilution
Procedure
Calculation of the measured value
1:1
Equal volumes of sample + distilled water
Read value × 2 = °dH
1:4
1 part sample + 4 parts distilled water
Read value × 5 = °dH
Specifications at a glance
Property
Value
Parameters
Carbonate hardness (CH)
Item number
31-075200
Method
Acid-base titration (syringe method)
Measuring principle
Syringe titration with color change
Measuring range
0 to 20 °dH
Resolution
Stepless (at least 0.5 °dH)
Analysis duration
1 minute
Number of provisions
approx. 200
Packaging
Plastic box
weight
approx. 250 g
Scope of delivery
1 precision dosing syringe with scale
1 test tube
2 reagent bottles (60 ml total)
Detailed operating instructions (DE/EN)
Safety data sheets
Robust plastic box
Reagent details
Reagent
Volume
Shelf life
Notes
Titration solution 1
60 ml
60 months (60M)
Shake before use
Titration solution 2
60 ml
60 months (60M)
Shake before use
Shelf life: The shelf life starts from the date of manufacture (printed on the bottle).
After opening, the reagents should be used within 18 months.
Store the test kit in a dry, cool place (10-25°C) and away from direct sunlight.