Reagent for Carbonate hardness C25 series:
See also:
Reagent carbonate values C25 series
What is the carbonate value measured for?
The carbonate value (also known as the temporal value) describes the proportion of the water value that is caused by hydrogen carbonate ions (HCO₃-). Monitoring carbonate levels is crucial in many areas of water treatment:
Cooling water systems: The carbonate values directly influence the lime-carbonic acid balance. Excessive values lead to lime precipitation in cooling towers and heat exchangers.
Boiler feed water: Hydrogen carbonate decomposes to carbonate and CO₂ during heating, which leads to corrosion in the steam/condensate system.
Swimming pool and bathing water: The carbonate value is an important parameter for the water balance according to Langelier or the LSI index.
Drinking water: The carbonate content influences the taste and the tendency for limescale formation in pipes.
How does the Reagent work?
The C25 Reagents are liquid color reagents for fully automatic water analysis with SYCON analyzers. The SYCON device automatically takes a water sample, precisely doses the Reagent and measures the resulting coloration photometrically.
Depending on the device mode, the system operates as a limit value monitor (SYCON 2602) or as a measuring range device (SYCON 2702, 2802) with continuous measured value output.
Advantages of the C25 series
4 concentration levels from 0.3 to 22.5 °dKH
Customized selection for different applications
Limit value and measuring range operation possible
Flexible use with SYCON 2602, 2702 and 2802
Fully automatic measurement around the clock
Leak-free monitoring without manual intervention
High-quality raw materials and our own recipe
Manufactured by RLS Wacon – Made in Germany
Single batch calibration
Consistently high precision with every delivery
24 months shelf life, 500 ml and 4-pack available
Economical container sizes for every need
Areas of application
Cooling water
Boiler feed water
Swimming pool / bathing water
Drinking water treatment
Industrial water treatment
Softening systems
Changing the indicator on the device
Only use indicator bottles with a capacity of 500 ml. They can be changed directly on the SYCON analyzer without tools.
Procedure
⚠ Wear protective gloves! Avoid skin contact with the reagent at all costs.
Deactivate automatic mode on the device.
Loosen the screw cap on the bottle and insert a new indicator bottle – collect any drips.
Flush the measuring chamber (menu > Service > Manual flushing).
Pump the Reagent into the measuring chamber (menu > Service > Pump Reagent) until the Reagent arrives free of bubbles. Confirm with [OK].
Indicates whether a full bottle has been inserted (fill level indicator).
Rinse the measuring chamber again and test the analysis procedure by starting an analysis.
Activate automatic mode again.
Important notes
Check the expiration date on the bottle before use – only use Reagent that has not expired.
When changing to a different measuring range, the device must be reconfigured accordingly.
If you have any questions about the configuration, please contact our technical support.
Measuring range
Reagent
Limit value (°dKH)
Measuring range (°dKH)
Article no. (1x 500 ml)
Item no. (4x 500 ml)
C25-1
1
0,30 – 7,5
32-086125
32-486125
C25-1,5
1,5
0,45 – 11,5
32-086135
32-486135
C25-2
2
0,60 – 15,0
32-086145
32-486145
C25-3
3
0,90 – 22,5
32-086155
32-486155
Unit conversion:
The measuring ranges are given in °dKH. Conversion factors: 1 °dKH = 0.3566 mmol/l = 1.7848 °f (French degrees of hardness).
What to do if measured values are out of range?
Measured value above the measuring range
Change to an Reagent with a higher limit value (e.g. from C25-1 to C25-2). The device must be reconfigured accordingly.
Measured value below the measuring range
Change to a Reagent with a lower limit for more precise results in the low range.
Specifications at a glance
Property
Value
Parameters
Carbonate value (temporal hardness)
Series
C25
Method
Colorimetric (photometric evaluation)
Measuring principle
Fully automatic Reagent dosing and color measurement
Base unit
°dKH
Available limit values
1 / 1.5 / 2 / 3 °dKH
Number of variants
4 concentration levels
Container size
500 ml single bottle
Value pack
4 × 500 ml
Shelf life
24 months from date of manufacture
manufacturer
RLS Wacon analytics GmbH, Hildesheim
Compatible devices
Device
Operating mode
Status
SYCON 2602
Limit value
Active
SYCON 2702
Measuring range
Active
SYCON 2802
Measuring range
Active
SYCON 2501
Limit value + measuring range
Discontinued
SYCON 2502
Limit value + measuring range
Terminated
SYCON 2800
Measuring range
Discontinued
Storage conditions
Store in a cool, dry and dark place. Do not freeze. Protect from direct sunlight. Shelf life: 24 months from date of manufacture (see label on the bottle). Use as soon as possible after opening.
Quality control
All Reagents are manufactured in-house at RLS Wacon according to strict quality standards. Each batch is individually calibrated and tested – Made in Germany.