Reagent for Silicate:
See also:
Reagent silicate
What is silicate measured for?
Silicate (SiO₂) is a dissolved ingredient that occurs in almost all natural waters. Monitoring the silicate concentration is particularly important in the following areas:
Power plants and steam generation: At high temperatures and pressures, silicate forms extremely hard, glass-like deposits on turbine blades and in steam pipes. These deposits are almost impossible to remove and lead to considerable efficiency losses.
Boiler feed water: Even low silicate concentrations can cause damage in the high-pressure range. Typical limit values are in the range of a few µg/l.
Reverse osmosis systems: Silicate can irreversibly foul membranes (silica fouling). Monitoring the permeate protects against costly membrane damage.
Mixed bed ion exchanger: Control of the exchange capacity – silicate is often one of the first parameters to break through during depletion.
How does the Reagent work?
The silicate Reagents consist of a reagent set with several components for fully automatic silicate determination with the SYCON SiO₂ analyzer. The device automatically takes a water sample, adds the reagents in the correct order and quantity and measures the coloration photometrically.
Two measuring ranges are available: Low Range (0.02-10 mg/l) for the trace range and High Range (10-150 mg/l) for higher concentrations.
Advantages of silicate Reagents
Two measuring ranges: Low Range (0.02-10 mg/l) and High Range (10-150 mg/l)
Covered from trace levels to high concentrations
Fully automatic measurement with the SYCON SiO₂
Continuous 24/7 monitoring of critical processes
Reagent sets with coordinated components
Optimal reaction conditions for precise results
High-quality raw materials and our own recipe
Manufactured by RLS Wacon – Made in Germany
Single batch calibration
Consistently high precision with every delivery
24 months shelf life
Cost-efficient stockpiling
Areas of application
Power plants / steam generation
Boiler feed water
Reverse osmosis systems
Mixed bed ion exchanger
Semiconductor industry / ultrapure water
Industrial water treatment
Changing reagents on the device
Several reagents are required for the silicate measurement. All bottles are connected to the SYCON SiO₂ at the same time. Only use bottles with a capacity of 500 ml.
Procedure
⚠ Wear protective gloves! Avoid skin contact with the reagents at all costs.
Deactivate automatic mode on the device.
Loosen the screw caps on the bottles and insert new reagent bottles – absorb any drips.
Flush the measuring chamber (menu > Service > Manual flushing).
Pump the Reagent into the measuring chamber (menu > Service > Pump Reagent) until all reagents arrive bubble-free. Confirm with [OK].
Indication of whether full bottles have been inserted (fill level indicator).
Rinse the measuring chamber again and test the analysis procedure by starting an analysis.
Activate automatic mode again.
Important notes
Check the expiration date on the bottles before use – only use reagents that have not expired.
When switching between Low Range and High Range, the device must be reconfigured accordingly.
Low Range requires 3 reagents (A, B, C), High Range only 2 (B, C).
If you have any questions about the configuration, please contact our technical support.
Measuring range
Reagent
Measuring range (mg/l SiO₂)
Item no. (set)
Composition (4-pack)
Low Range
0,02 – 10
32-010100
2x A, 1x B, 1x C
High Range
10 – 150
32-010110
2x B, 2x C
Reagent composition:
Low Range requires reagents A, B and C. High Range requires reagent B and C only. Single bottles: A = 32-010101, B = 32-010102, C = 32-010103.
What to do if measured values are out of range?
Measured value > 10 mg/l (when using Low Range)
Change to the High Range reagent set for the extended measuring range up to 150 mg/l. The device must be reconfigured accordingly.
Measured value < 10 mg/l (when using High Range)
Change to the Low Range reagent set for more precise measurements in the trace range.
Specifications at a glance
Property
Value
Parameters
Silicate (SiO₂)
Method
Colorimetric (photometric evaluation)
Measuring principle
Fully automatic multi-reagent dosing and color measurement
Unit
mg/l SiO₂
Measuring range Low Range
0.02 – 10 mg/l
Measuring range High Range
10 – 150 mg/l
Number of reagents
LR: 3 (A, B, C) / HR: 2 (B, C)
Container size
500 ml single bottles
Shelf life
24 months from date of manufacture
manufacturer
RLS Wacon analytics GmbH, Hildesheim
Scope of delivery of the sets
set
Item no.
Contents
Low Range Set
32-010100
2× reagent A (500 ml), 1× reagent B (500 ml), 1× reagent C (500 ml)
High Range Set
32-010110
2× reagent B (500 ml), 2× reagent C (500 ml)
Single bottles
Reagent
Item no.
Volume
Reagent A
32-010101
500 ml
Reagent B
32-010102
500 ml
Reagent C
32-010103
500 ml
Compatible devices
Device
Operating mode
Status
SYCON SiO₂
Measuring range
Active
Storage conditions
Store in a cool, dry and dark place. Do not freeze. Protect from direct sunlight. Shelf life: 24 months from date of manufacture (see label on the bottle). Use as soon as possible after opening.
Quality control
All reagents are manufactured in-house at RLS Wacon according to strict quality standards. Each batch is individually calibrated and tested – Made in Germany.