Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 --- Our telephone system is currently out of service due to line work. You can reach us by e-mail at info@rls-wacon.de or by calling +49 (0)151/1054 1979 ---

Sample cooler AC 100 with valves

Article number: 
Category
Quantity
Add to Quote
Reagent for Sample cooler AC 100 with valves:

The PC 200 pre-cooler is used to take a water sample from a hot water pipe. By installing it in the sampling line and connecting a cold water line, samples such as boiler water, boiler feed water and condensate can be automatically monitored for quality. Our SYCON BOB online analyzer is used downstream of the sample cooler to determine the water hardness.

The special features of the PC series sample coolers include

  • Compact design
  • Optimally matched to the analyzers of the Sycon series
  • Use of high-quality stainless steel
  • Sample water inlet pressure up to 16 bar
  • Inlet pressure cooling water up to 8 bar

The AC 100 sample cooler is used to take a water sample for manual analysis from a hot water pipe. By installing it in the sampling line and connecting a cold water line, samples such as boiler water, boiler feed water and condensate can be qualitatively monitored. The samples are then analyzed using our CALDUR test kits, for example.

The special features of the PC series sample coolers include

  • Compact design
  • Use of high-quality stainless steel
  • Inlet pressure cooling water up to 8 bar

Parameters Value / Range
Weight 2.4 kg without valves / 2.72 kg with valves
Dimensions Body 100 x 225 mm / cooling coil 2.6 m x 8 mm
Material Stainless steel 1.4571, electropolished

Parameters Value / Range
Installation Wall mounting in closed rooms
Cooling water inlet 8 mm
Cooling water outlet 1/2″ external thread
Sample water input 8 mm
Sample water output 8 mm
Nominal pressure 8 bar (open outlets)
Cooling coil 16 bar
Fastening 2-fold, spacing 54 mm, screw diameter 4 mm

Item number Article description
30-012110 Hot water valve type HWV 100, fine control valve made of brass, PN 50 bar, 100°C
30-012121 Cold water valve type KWV 150, 1/4ball valve nickel-plated brass with screw connections for 8 mm pipe