The SYCON Fe is an analyzer for measuring free iron in a water sample. In addition to the usual relays and input contacts, the online analyzer has a 20 mA current and RS485 interface with Modbus RTU protocol.
Iron (Fe) is one of the most common interference parameters in water treatment. Even low concentrations lead to visible discoloration of the water, rusty-brown deposits in pipes and can damage downstream treatment stages such as softening systems or reverse osmosis membranes. A limit value of 0.2 mg/l applies in the drinking water supply – exceeding this value leads to complaints and indicates corrosion or inadequate treatment.
In well water treatment, iron is often the determining parameter for the design of the deferrization system. Continuous online monitoring detects concentration fluctuations in the raw water and checks the efficiency of the treatment plant.
The SYCON Fe from RLS Wacon is a fully automatic online analyzer for the quantitative determination of the iron concentration in process water. It works according to the colorimetric method and covers a measuring range of 0.01 to 6 mg/l with two Reagents (Fe LRS and Fe HRS).
The device offers four relay outputs, an RS485 interface with Modbus RTU, an analog output 0/4-20 mA and two digital inputs. In addition to the programmable limit value alarm, the SYCON Fe has a message for measured values below the detection limit – an important signal for monitoring deferrization systems.
The SYCON Fe works according to the colorimetric method. For each analysis, a water sample is automatically fed into the measuring chamber and mixed with the Reagent. The reaction with dissolved iron produces a characteristic color reaction whose intensity is proportional to the iron concentration. An optical sensor evaluates the color intensity and calculates the measured value.
Depending on the indicator selected, a corresponding actuator LED is used in the measuring chamber: blue marker for the Reagent Fe LRS (low concentrations) and red marker for Fe HRS (higher concentrations). The analysis time is approx. 7 minutes.
The Fe LRS Reagent already detects concentrations far below the drinking water limit of 0.2 mg/l
Fe LRS (0.01-0.5 mg/l) for drinking water and treated water, Fe HRS (0.2-6 mg/l) for untreated water and industrial applications
Confirms complete removal of iron – important for monitoring deferrization systems
Seamless integration into PLC and SCADA systems for central process monitoring
Limit value alarm, detection limit / dosing fault, device fault, analysis active / cooling water valve / indicator message
Direct transmission of measured values to the system controller
Automatic compensation of ambient light and soiling
Same dimensions as SYCON 2702/2802 – as wall bracket (IP43) or in housing (IP54)
Complete logging, installation wizard, software updates via SD card
The SYCON Fe is used wherever the iron content in water needs to be continuously monitored:
| Area of application | Description |
|---|---|
| Drinking water treatment | Monitoring the iron content after iron removal – the drinking water limit is 0.2 mg/l. Even slight exceedances lead to visible discoloration and customer complaints |
| Well water treatment | Continuous monitoring of the raw water to assess the iron load and control the efficiency of the deferrization system |
| Corrosion control | Elevated iron levels in mains water indicate corrosion in pipes – early detection prevents major damage |
| Industrial water treatment | Protection of downstream system components (membranes, ion exchangers, heat exchangers) against iron-related deposits and fouling |
| Parameters | Requirement |
|---|---|
| Texture | Clear, free of solids, free of gas bubbles |
| Temperature | 5-40 °C (use sample cooler for higher temperatures) |
| min. 0.5 bar, max. 5 bar (recommended 1-2 bar) |
Due to the measuring method, the SYCON Fe requires shorter cleaning intervals than the SYCON series hardening devices:
| Interval | Measure |
|---|---|
| Every 3 months | Cleaning the measuring chamber (with SYCON Clean cleaning set incl. FIT 3000) |
| Every 15,000 analyses or 24 months | Installation of the maintenance set: Replacement of the peristaltic pump cassette, seals and agitator blades |
Bright light sources with fluctuating brightness in the immediate vicinity of the measuring chamber have an unfavorable effect on the measuring accuracy and should be avoided. A sample cooler is required for sample water temperatures above 40 °C.
| Reagent | Measuring range (mg/l) | Actuator LED | Item no. |
|---|---|---|---|
| Fe LRS | 0,01 – 0,5 | Blue marking | 32-080320 |
| Fe HRS | 0,2 – 6 | Red marking | 32-080330 |
When changing between the Reagents, the actuator LED in the measuring chamber must also be replaced.
| Property | Value |
|---|---|
| Measuring principle | Colorimetric method |
| Parameters | Dissolved iron (Fe) |
| Measuring range Fe LRS | 0.01-0.5 mg/l (ppm) |
| Measuring range Fe HRS | 0.2-6 mg/l (ppm) |
| Accuracy | ± 10 % of the maximum measured value |
| Repeatability | ± 5 % |
| Duration of analysis | approx. 7 minutes |
| Indicator consumption | approx. 0.5 ml per analysis |
| Analysis per 500 ml bottle | approx. 1.000 |
| Water consumption | approx. 2 l per analysis at 2 bar |
| Durability Reagent | min. 1 year (at < 25 °C, stored in the dark) |
| Property | Value |
|---|---|
| Supply voltage (230 V version) | 100-240 V AC, 50/60 Hz |
| Supply voltage (24 V version) | 21.6-26.4 V AC/DC, 50 Hz |
| Power consumption | max. 25 VA / 8 VA measurement mode / 3.5 VA standby |
| Relay outputs | 4 relays, max. 250 V AC/DC, 1 A, potential-free changeover contacts (NC/NO) |
| Relay 1 | Limit value alarm |
| Relay 2 | Below the detection limit / dosing fault |
| Relay 3 | Device fault |
| Relay 4 | Analysis active / cooling water valve / indicator message |
| Analog output | 0-20 mA or 4-20 mA, resolution < 100 µA, max. 750 Ω Load |
| RS485 interface | Modbus RTU |
| Digital inputs | 2× galvanically isolated – analysis start, water meter, flow monitor, interval reset |
| Feature | Wall bracket IP43 | Housing IP54 |
|---|---|---|
| Dimensions (W × H × D) | 274 × 275 × 129 mm | 332 × 345 × 190 mm |
| weight | approx. 2.0 kg | approx. 4.0 kg |
| Protection class | IP43 | IP54 |
| Ambient condition | Value |
|---|---|
| Operating temperature | 10-45 °C |
| Storage temperature | 0-45 °C |
| Sample water temperature | 5-40 °C |
| Air humidity | 20-90 % RH (non-condensing) |
| Inlet water pressure | min. 0.5 bar, max. 5 bar (recommended 1-2 bar) |
| Reagent | Measuring range | Actuator LED | Item no. |
|---|---|---|---|
| Fe LRS | 0.01-0.5 mg/l | Blue marking | 32-080320 |
| Fe HRS | 0.2-6 mg/l | Red marking | 32-080330 |
Note: The measured value to be monitored should be in the middle of the Reagent measuring range if possible. When changing the indicator between LRS and HRS, the actuator LED in the measuring chamber must also be replaced.
| Property | Value |
|---|---|
| Display | Multicolored graphic display |
| Operation | 8 buttons |
| Languages | German, English, French, Italian, Spanish, Russian, Polish, Dutch, Danish |
| Data memory | SD card (measured value logging and software updates) |
| Component |
|---|
| 1 × SYCON Fe (configured for selected supply voltage) |
| Wall bracket or housing (depending on order variant) |
| SD card |
| Operating instructions (German or English, depending on order variant) |
Note: A compatible Reagent (Fe LRS or Fe HRS) with matching actuator LED is required for operation and must be ordered separately.
| Design | |
|---|---|
| Power supply | 230 V AC, 24 V AC/DC |
| Manual | English, German |
| Parameter | Iron |
| Product category | Analyzer |
| Series | SYCON |