Maintenance set for SYCON analyzers – Reliable operation through regular maintenance
Why is regular maintenance important?
The SYCON analyzers from RLS Wacon work fully automatically and reliably – provided that the wearing parts are replaced at regular intervals. The peristaltic pump cassette is the central dosing element: it delivers the Reagent precisely into the measuring cell and thus significantly determines the measuring accuracy. Worn pump hoses lead to inaccurate dosing quantities and therefore to falsified measurement results.
Seals (O-rings), suction lance and agitator blades are also subject to natural wear due to constant contact with water and indicator solution. Replacing these components in good time prevents leaks, dosing errors and unplanned device failures.
When should you wait?
The maintenance intervals vary depending on the appliance type:
Device type
Cleaning the measuring chamber
Installation of the maintenance set
SYCON 2502/2602
Every 6 months
Every 30,000 analyses or 24 months
SYCON 2702/2802/P
Every 6 months
Every 30,000 analyses or 24 months
SYCON Fe
Every 3 months
Every 15,000 analyses or 24 months
SYCON SiO2
Every 4 weeks
Every 10,000 analyses or 6 months
SYCON BoB
Every 6 months
Every 15,000 analyses or 12 months
If the water or ambient temperature is high or the water is organically contaminated, the cleaning intervals should be shortened if necessary. Cleaning the measuring chamber takes approx. 20 minutes.
What is included in the maintenance kit?
Each maintenance set contains all the wear parts required for complete maintenance: Hose pump cassette, bottle connector, suction lance, O-rings, dosing O-ring and agitator blade. The exact scope is tailored to the respective appliance type.
Advantages of the original maintenance set
All wear parts in one set
No tedious searching for individual spare parts – everything is coordinated
Original RLS Wacon quality
Precisely manufactured components for maximum measuring accuracy and service life
Perfect fit for your device type
Select the set for your specific SYCON model – with the right pump cassettes and seals
Easy replacement without special tools
All components can be replaced in a few simple steps – detailed instructions in the device manual
Plannable maintenance costs
One set covers complete maintenance – no hidden costs for missing parts
Maintenance instructions
The following instructions describe the replacement of wearing parts as part of regular maintenance. Detailed, device-specific instructions can be found in the respective operating instructions.
Safety instructions
Reagents, Reagents and cleaning fluids may contain hazardous chemical components. Observe the relevant safety data sheets (SDS) before starting work.
Always wear protective goggles and gloves during maintenance.
Avoid skin contact with reagents and cleaning agents. In case of contact, rinse immediately with water.
Only carry out maintenance work when the appliance is switched off and depressurized.
Before maintenance
Wait for ongoing analysis – never intervene during an active measurement
Set the device to maintenance mode via the menu or disconnect the power supply
Shut off the water supply and release residual pressure
Remove the reagent or indicator bottle(s) and pull out the suction lance(s)
Replacing the components
Component
Procedure
Peristaltic pump cassette
Remove the cassette from the holder, loosen the hose ends. Insert new cassette and connect hoses as marked. Ensure that the correct type of cassette is used (white or black rollers depending on the appliance model).
O-rings
Remove old O-rings from the measuring chamber and the plug. Clean the sealing surfaces. Rub new O-rings lightly with acid-free grease and insert. Note: Dosing O-rings must not be greased.
Suction lance
Remove the old suction lance and replace it with the new one. Make sure that the lance reaches the bottom of the bottle so that the reagent can be fully utilized.
Bottle connector
Remove the old bottle connector and replace it with the new one. Check that it is firmly attached to the reagent bottle.
Agitator blade
Replace the stirring paddle in the measuring chamber. The stirring paddle ensures uniform mixing of the sample and reagent during the analysis.
Special feature SYCON SiO2
The SYCON SiO2 works with a 3-reagent system (A, B and C). During maintenance, all three peristaltic pump cassettes must therefore be replaced at the same time (1× with black rollers for channel A, 2× with white rollers for channels B and C). Three suction lances, three bottle connectors and three dosing O-rings must also be replaced. In addition, the screw-on dispensing plugs must be loosened and the light rod plug carefully removed (sensitive to scratches – only push out with a plastic or wooden stick). After reassembly, all three reagents must be added individually until no more air bubbles escape.
After maintenance
Check all connections for tight fit and leaks
Insert or reconnect reagent or indicator bottle(s)
Open the water supply and check for leaks
Switch on the device and add reagent/reagent until there are no more air bubbles in the tubes
Perform test analysis and check measurement result with known reference or manual test set (e.g. CALDUR®)
Reset the maintenance counter in the device menu
Additional cleaning with SYCON Clean
In addition to replacing the wearing parts, RLS Wacon recommends cleaning the measuring chamber regularly with the SYCON Clean cleaning set (item no. 30-010900). The cleaning set removes deposits and soiling in the measuring chamber and the lines, which can impair the measuring accuracy. The device-specific cleaning intervals can be found in the maintenance table in the product description.
Tip for maintenance planning
Order the maintenance set in good time before the planned maintenance date. The SYCON devices log the analysis counter on the SD card – this allows the optimum maintenance time to be determined based on actual usage.
Important notes
Only use original maintenance kits from RLS Wacon – third-party components can impair the measuring accuracy and tightness
For SYCON 2502/2602: Use hose pump cassette with black rollers – the cassette with white rollers is not suitable for these appliances
For SYCON SiO2: Replace all three pump cassettes at the same time, correctly assign reagent bottles A, B and C
Perform a test analysis after each maintenance and compare the result with a reference
Dispose of used O-rings, pump cassettes and reagent residues in accordance with local regulations