SYCON P – Online titration device for automatic measurement of the p-value (alkalinity), RS485 Modbus RTU, 4 relay outputs, 2 limit values, analog output 4-20 mA
SYCON P – Online titration device for automatic measurement of the p-value
What is the p-value and why is it measured?
The p-value (phenolphthalein alkalinity) describes the buffering capacity of water against acid input. It measures the hydroxide ions (OH-) and carbonate ions (CO₃²-) dissolved in the water, which are converted into water (H₂O) and hydrogen carbonate (HCO₃-) during titration. The measurement result is given in mmol/l.
Without a sufficient p-value, water has hardly any buffering capacity – even small amounts of acid can shift the pH value into the acidic range. In practice, the p-value is therefore a decisive parameter for corrosion protection: in steam boiler systems, cooling water circuits and industrial treatment plants, the alkalinity of the water must be continuously monitored in order to prevent corrosion on pipes, heat exchangers and boiler components.
What is the SYCON P?
The SYCON P from RLS Wacon is a fully automatic online titration device for the quantitative determination of the p-value in process water. It is based on the proven SYCON platform and offers the full range of functions of the Premium series: two programmable limit values, four relay outputs, an RS485 interface with Modbus RTU and an analog output 0/4-20 mA.
With two Reagents (P25-5 and P25-20), the device covers a measuring range of 0.2 to 20 mmol/l and is therefore suitable both for the fine monitoring of low-alkaline water and for highly alkaline applications.
How does the SYCON P work?
The SYCON P works according to the colorimetric titration method. The dye phenolphthalein contained in the Reagent is colorless below pH 8.2 and intensely pink to violet above this level. During the measurement, an alkaline water sample (pH > 8.2) is automatically fed into the measuring chamber and titrated step by step with the Reagent. An optical sensor detects the color change at the pH 8.2 point and calculates the p-value.
The analysis time is approx. 3 minutes. The indicator consumption is only 0.05 to 0.6 ml per analysis, depending on the measured p-value.
Advantages of the SYCON P
Exact determination of measured values by titration method
Quantitative determination of the p-value with ± 5 % accuracy and ± 2.5 % repeatability
Continuous online monitoring of a critical corrosion parameter
Protects boilers, pipes and heat exchangers from acid-induced corrosion
Two measuring ranges for different applications
P25-5 (0.2-5 mmol/l or 0.01-0.5 ppm) for low alkaline water, P25-20 (0.8-20 mmol/l or 0.2-6.0 ppm) for highly alkaline applications
RS485 interface with Modbus RTU
Direct query of measured values and device parameters by a PLC – seamless integration into the system automation
Two freely programmable limit values
Pre-alarm and main alarm or monitoring of a bandwidth with lower and upper threshold value
4 potential-free relay outputs
Limit value 1, limit value 2 / indicator level, device fault, analysis active / cooling water valve
Full color sensor with automatic zero adjustment
Zero sample before every measurement – automatic compensation for ambient light, contamination and turbidity, no manual calibration needed
Very low indicator consumption
0.05-0.6 ml per analysis – one 500 ml bottle is sufficient for numerous analyses
9-language graphic display with SD card
Complete logging, installation wizard, software updates via SD card
The SYCON P is a titration device for measuring the +p value. In addition to the usual relays and input contacts, the online analyzer has a 20 mA current and RS485 interface with Modbus RTU protocol.
Field of application
The SYCON P is used wherever the alkalinity of the process water needs to be continuously monitored:
Area of application
Description
Steam boiler systems
Monitoring the alkalinity in the boiler water to protect against acid corrosion – too low a p-value endangers boiler tubes and heat exchangers
Cooling water circuits
Continuous monitoring of the buffer capacity in the circulating water – basis for the correct dosing of conditioning chemicals
Industrial process water treatment
Quantitative p-value monitoring with automatic documentation and PLC connection via Modbus RTU or 4-20 mA
Condensate return
Monitoring the returning condensate for sufficient alkalinity – a drop in acidity in the condensate indicates leaks in the heat exchanger
Requirements for the sample water
Parameters
Requirement
pH value
> 8,2
Characteristics
Clear, colorless, free of solids, free of gas bubbles
Temperature
5-40 °C (use sample cooler for higher temperatures)
Print
min. 0.5 bar, max. 5 bar (recommended 1-2 bar)
Installation and commissioning
Mount the appliance on a dry, easily accessible wall – upright, with free access to the drain
Make water connections: Inlet and outlet via 6 mm hose
Electrical connections via the cable glands: Supply voltage, 4 relay outputs, analog output, RS485, 2 digital inputs
Insert the Reagent of the P25 series and insert the suction lance
Start installation wizard – step-by-step guide through all settings including both limit values
Start the first analysis run and check the measured value
Tip for the installation
A sample cooler is required for sample water temperatures above 40 °C. Relay 4 can be configured to automatically activate a cooling water valve. Oxidizing agents (chlorine, ozone) above the drinking water limit values interfere with the measurement – in this case, an upstream activated carbon filter is recommended.
Important notes
Only for indoor use and wall mounting in upright position
The sample water must have a pH value above 8.2 – no p-value measurement is possible below this value
A compatible P25 Reagent is required for operation
Free, short drain without back pressure required
Install maintenance set (item no. 33-090034) every 2 years or 30,000 analyzes
Clean the measuring chamber every 6 months – use SYCON Clean (item no. 30-010900)
Measuring range and Reagents
Reagent
Measuring range (mmol/l)
Measuring range (ppm)
Item no. (single bottle)
Item no. (4-pack)
P25-5
0,2 – 5
0,01 – 0,5
32-086335
32-486335
P25-20
0,8 – 20
0,2 – 6,0
32-086345
32-486345
The measured value to be monitored should be in the middle of the Reagent measuring range if possible.
Specifications at a glance
Property
Value
Measuring principle
Colorimetric titration method
Parameter
p-value (phenolphthalein alkalinity)
Measuring range P25-5
0.2-5 mmol/l (0.01-0.5 ppm)
Measuring range P25-20
0.8-20 mmol/l (0.2-6.0 ppm)
Accuracy
± 5 % of the maximum measured value
Repeat accuracy
± 2,5 %
Unit
mmol/l
Analysis time
approx. 3 minutes
Indicator consumption
0.05-0.6 ml per analysis (p-value dependent)
Water consumption
approx. 1 l per analysis at 2 bar
Durability Reagent
at least 12 months (stored at < 25 °C, in the dark)
Sample water requirement
pH > 8.2, clear, free of solids, without gas bubbles
Electrical data
Property
Value
Supply voltage (230 V version)
100-240 V AC, 50/60 Hz
Supply voltage (24 V version)
21.6-26.4 V AC/DC, 50 Hz
Power consumption
max. 25 VA / 8 VA measurement mode / 3.5 VA standby
Relay outputs
4 relays, max. 250 V AC/DC, 1 A, potential-free changeover contacts (NC/NO)
Relay 1
Limit value exceeded 1
Relay 2
Limit value exceeded 2 / indicator level low
Relay 3
Device fault
Relay 4
Analysis active / cooling water valve / indicator level low
Analog output
0-20 mA or 4-20 mA, resolution < 100 µA, max. 750 Ω Load